Turbo
Molecular Pumps
YTP Series Turbo Molecular Pump
Evacuation Units
The YTP Series are simple high-vacuum
evacuation units that each consist of an ULVAC turbo molecular
pump (UTM Series pump), rotary pump, forevalve, Pirani vacuum
gauge and control system. A single unit can evacuate from
atmospheric pressure down to high vacuums.
Features
- Two types are available : Automatic operation
type (YTP-SA Series) and manual operation type (YTP-M Series).
In addition, 4 evacuation speeds can be selected.
- During power failures, the forevalve closes
automatically, and the atmosphere in the rotary pump is
vented automatically.
The evacuation unit doesn't operate when the power is restored.
- The turbo molecular pump uses fluorine-based
oil with a saturation vapor pressure of around 10-11
Pa (at 20ºC), so the atmosphere stays completely oil-free.
- The turbo molecular pump uses high-performance
bearings, so doesn't require time-consuming oil changes
or bearing replacements.
- YTP-SA models come with interlock functions
and external control functions, and can be upgraded with
a wide array of functions.
Applications
- As evacuation systems for analysis equipment
- As rough vacuum pumps for accelerators and
beam lines
- As high-vacuum evacuation systems for laboratory
equipment
Specifications
| Model |
YTP- 50M/SA |
YTP- 150M/SA |
YTP- 300M/SA |
YTP- 500M/SA *1 |
| TMP inlet flange size |
CF |
114 |
152 |
203 |
| JIS/VG *1 |
65 |
100 |
150 |
| TMP ultimate pressure (Pa) (*2,
*3) |
10-7 order |
10-8 order |
TMP pumping speed (l/s)
(*3, *4) |
N2 |
50 |
190 |
300 |
550 |
| TMP bearing type |
Top : Permanent magnet bearings ; Bottom
: Pivot bearings |
| RP ultimate pressure (Pa) |
6.7 x 10-2 |
| RP pumping speed (l/min) |
50Hz |
50 |
100 |
200 |
| 60Hz |
60 |
120 |
240 |
| TMP back-pressure/rough evacuation
monitor |
Pirani vacuum gauge (10-1 to
2.7 x 103Pa) |
| Input voltage |
Single-phase AC100V(50/60Hz) |
| *1 |
Manufactured to order |
| *2 |
Value after bake-out of turbo molecular pump alone. |
| *3 |
Value measured in accordance with JVIS005 standard,
or calculated value based on it |
| *4 |
Evacuation speed values are values when protective
mesh isn't used. |