Ulvac Home

    

Search:

Contact Us
 


   ULVAC Technologies, Inc.
   401 Griffin Brook Drive
   Methuen, MA 01844

   Tel: 978-686-7550
   Fax: 978-689-6300

 Residual Gas Analyzer 

Qulee

"Qulee" (pronounced as "KuLee"), is the Japanese word for a 4-leaf clover, and is the name for ULVAC's newest quadrupole residual gas analyzer.  This CGM (Compact Gas Monitor) can measure residual gas pressures during processes and achieves high measurement accuracy and resolution up to a high pressure of 1 Pa without the need for differential pumping.  This unit also features the utmost in simplicity of design and operation.  Many requests from facility engineers working in production lines across a broad range of industries have been incorporated into the features of this new RGA.

Features

  • High resolution quadrupole mass spectrometer
  • High accuracy at high pressure (1Pa) without differential pumping
  • One touch operation
  • Displays partial pressure of He/H20/N2/O2/Any Gas
  • Total pressure function
  • Bakeable to 250°C
  • Mass range 1-50 amu

Applications

  • Residual gas analysis in PVD system
  • Gas monitoring of PVD system during processing

 

Specifications

Item
Qulee CGM
Model No.
CGM-051
CGM-052
Mass Filter Type
Quadrupole
Mass Range
1 - 50 amu
Resolution
M/ M=1M(10% P.H.)
Sensitivity
0.1uA/Pa
0.1mA/Pa
Minimum Detectable Partial Pressure
1X10-7Pa
1X10-10 Pa
Detector Type
Faraday Cup
EM Tube/Faraday Cup
Maximum Operating Pressure
2Pa
2Pa(FC)
1e-2Pa(SEM)
Linearity
1Pa
Total Pressure Function
capable
One Click function
capable/He/H2O/N2/O2/Any Gas
One Click Leak Test function
capable
Power Input
DC24V
Maximum Sensor Operating Temperature
120°C
Maximum Bakeout Temperature(electronics removed)
250°C
Operating Temperature
10-40°C
Standard Software
Qulee QCS (Windoes2000/XP)

 

Applications

This “Qulee QCS” is a software for Quadruple Mass Spectrometer that can be used for almost all models of ULVAC Mass Spectrometer. RS232C and RS485 are applicable. It enables you to control the units and convert the data to CSV easily. Our old model, that includes QCS 2001 or MSQ-400, MM200 etc., can upgrade to “Qulee QCS”. Please contact us for detail.

 


Features

  • Possible to connect up to 8 sets
  • Recipe setting
  • Easy operation by short-cut button
  • Easy leak check in the helium leak test mode
  • Includes background subtraction and measured area
    integration
  • Save 2 analog data at once
  • Possible to set 2 partial pressure set-point output
  • Possible to convert the data into CSV easily


Scan Mode

 

 

 

 

 

Trend Mode

 

 

Model
Qulee QMS
OS
Windows 98, Me, NT4, 2000, XP
Interface
RS-232C/485
Sensor Number
Max. 8
Sensor model
Qulee/REGA/REPROS/
SEPION (+SIM)/
ST-400 (+SIM*)
Recipe
100 (user to setup and control 70)
Sweep speed
50, 100, 200, 500, 1000, 2000 msec/mass
Sampling interval
Auto, 0.5 - 1000sec (Trend mode),
1 - 1000 sec (Scan mode)
Measurement end time
sequence or at random is available
Measurement mode
Scan mode:
Mode to scan arbitrary range continuously
Trend mode:
Mode to measure arbitrary mass number selectively
(20ch)
Leak-test mode:
One touch helion leak testing
Analog mode:
Mode to provide peak top setting value of spectrum
Sensitivity Calibration:
Mode to calibrate sensitive of SEM
Degas mode:
Mode to degas ion source (SEPION)
Indication of gas composition name
Able to indicate the gas composition name in scan and trend mode
Function
Background subtraction/Integration of measured area
Analog Input
2Ch: 0 - 10V
Set-point
2Ch (partial pressure set-point)
Ionization voltage
20 - 70eV (only REPROS)

 

Model
Qulee QCS
Process for error
Filament, SEM, RF, External
Display
Vertical axis :
Changeover between linear/logarithmic, selection of display range[A / Pa]
Horizontal axis :
Modifying the mass number and time and scrolling
Display of stored data
(Display during measurement is possible)
Setting of simultaneous display of a bar graph and
line graph is possible
Data Save
Auto save, confirmation, no save
The file name is generated from the clock of PC automatically
File can be converted to CSV format
Any desired file name can be set
Print Data
Printing of graph in scanning, trend, and analog mode is possible
Setting of the range of printing screen. Setting of colors of data and setting of the width of lines of trend graph
Option
RS-232C/485 converter, PC
Specification of PC
HDD has 2MB of empty capacity
(not including measurement data) More than 64MB
RAM and more, CPU is Pentium 200MHz and more Screen display area is 800 x 600 and more,
Floppy Drive is available.
COM port (RS232C) is available.
with CD-ROM drive



For more information
Please Email Us
TEL : 978-686-7550/ FAX : 978-689-6300

^top   
Copyright 2009. ULVAC Technologies, Inc. All rights reserved.| Terms for using this site .