Ulvac Home

    

Search:

Contact Us
 


   ULVAC Technologies, Inc.
   401 Griffin Brook Drive
   Methuen, MA 01844

   Tel: 978-686-7550
   Fax: 978-689-6300


Standard Laser Flash Thermal Constants Analyzer


TC9000TC 9000

ULVAC-RIKO systems make the difference in research and development and quality control of high-tech materials. ULVAC-RIKO's system use the widely accepted Laser Flash method to measure three thermal constants: thermal conductivity, thermal diffusivity, and specific heat. Since 1970, ULVAC-RIKO has delivered over 300 laser thermal constants measuring units around the world. Now ULVAC-RIKO presents its most advanced laser thermal constants measuring system: the TC-9000. ULVAC-RIKO's TC-9000 has the latest technology to meet the increasingly diversified needs for thermal measurement.



 

Specifications

                        Model
Item

TC-9000L

TC-9000H

TC-9000UHV

 

Measurement temperature

-150°C ~ 200°C

Room temperature to 1500°C

Room temperature to 2200°C

Sample Size

10mm dia. X 1mm to 3mm thick

Sample irradiating light

Glass laser output 10 J/pulse FWHM 0.1 msec/0.3 msec or less

Sensor

Thermocouple

Thermocouple, IR detector

IR detector

Measurement

Thermal diffusivity, specific heat capacity, thermal conductivity

Data analysis method

t ½ (half time) method, logarithmic method, curve fitting method*

Footprint

800 mm wide X 1050mm deep

1700 mm wide X 1500 mm deep

Power requirements

200 VAC single phase, 10A 100 VAC, 15 A each one

200 VAC single phase, 30A 100 VAC, 15 A each one

200 VAC single phase, 80A 100 VAC, 15 A each one

Cooling water requirements

City water 5L/min or more

Measurement atmosphere

Low pressure helium gas

Vacuum or air down to 150°C

Vacuum


Model
Item

TC9-Low

TC9-STEP

TC9-SB1

TC9-SB2

 

Measurement temperature

-150°C ~ 200°C

Room temperature to 1000°C

Room temperature

Room temperature to 200°C

Sample Size

Same as TC-9000

30mm dia X 3 to 5 mm thick

25mm square X 1mm or less thick

Sample irradiating light

Same as TC-9000

Halogen light

Same as TC-9000

Sensor

Thermocouple

Same as TC-9000

Measurement properties

Same as TC-9000

Thermal diffusivity

Data analysis method

Same as TC-9000

Step method

Same as TC-9000

Footprint

-

300 mm wide X 650 mm deep

-

Power requirements

-

100 VAC, 30 A

-

Cooling water requirements

City water 5L/min or more

-

Measurement atmosphere

Low pressure helium gas

Low pressure argon gas

Air

* The curve fitting method is the analysis method provided by the National Institute of Advanced Industrial Technology and Science

For more information
Please Email Us
TEL : 978-686-7550/ FAX : 978-689-6300
   
Copyright 2011. ULVAC Technologies, Inc. All rights reserved.| Terms for using this site .