Standard Laser Flash Thermal Constants Analyzer


TC7000

ULVAC-RIKO systems make the difference in research and development and quality control of high-tech materials. ULVAC-RIKO's system use the widely accepted Laser Flash method to measure three thermal constants: thermal conductivity, thermal diffusivity, and specific heat. Since 1970, ULVAC-RIKO has delivered over 300 laser thermal constants measuring units around the world. Now ULVAC-RIKO presents its most advanced laser thermal constants measuring system: the TC7000. ULVAC-RIKO's TC7000 has the latest technology to meet the increasingly diversified needs for thermal measurement.



Specifications

Standard Model
TC7000-HNC
Temperature Range
Room temperature to 1,500degC
Atmosphere

Vacuum Air or reducted pressure inert gas is also possible below 300°C

Sample Size
Standard mesurement method
10mm-dia. X 1 to 3mm thick disk solid sample.
For pure Aluminum sample, Thickness is 2mm minimum.
Back Temperature Sensor
Themocouple (R type) and infrared temperature sensor
Thermal Conductivity Measurement Range
0.1W to 400mW/mK
Measuring Reproducibility
Thermal diffusivity within +/- 3%
Specific heat within +/- 7%
Thermal conductivity within +/- 10%
Laser
Normal oscillation type glass laser (wavelength 0.69um)
Transient Memory Sampling Time
1us to 5ms/Word X 8K Words
Utility Requirement
Power: 200 or 220VAC, single phase, 40A
Cooling water: city water 7 lit/min.
Space Requirement
2,000mm wide X 1,300mm deep
Weight
About 350Kg
Option
Attachment for two dimensional method.
Sample size: Thin plate larger than 25mm square and
less than 1mm thick.