Leak Tester HELIOT model 102

Leak tests on items subject to internal pressure such as compressors and gas piping have conventionally been performed by methods such as the water immersion method or halogen test. However with the product quality improvements and production rationalization of recent years, users are demanding leak inspection equipment that provides more reliability and uses smaller and simpler system devices. HELIOT model 102 is a specialized sniffer leak tester that offers high-precision testing in a compact/space-saving, low-cost design.

Features

  • Use of helium gas for electrical measurement enables leak tests with high precision, stability and reproducibility.
  • Compact/lightweight design fits in palm of hand, and touch-panel display is easy to operate and read.
  • Dry pump type is available (comes with built-in membrane pump).
  • Highest level of response. Easy-to-operate suction probe is provided standard.
  • Tester is driven by DC power supply, for outstanding portability and easy startup in field service applications.

Applications

  • Quantitative tests replacing methods such as water immersion method : Automotive industry, air-conditioning/refrigeration industries, etc.
  • High-precision tests replacing air leak tests : Air-conditioning/refrigeration industries, electric appliance/electronics/semiconductor industries, etc.
  • High-precision/high-stability tests replacing acoustic method : Leak tests on underground pipes, etc.

Specifications

Model overview Sniffer type
Sensitivity 0.1-105ppm (Reference values 10-7 to 10-2Pa·m3/s) (*2)
Units Pa·m3/s, mbar·L/s, ppm
Main pump/fore pump Compound molecular pump/membrane pump
Sniffer pump Membrane pump
Analyzer tube Magnetic field deflection mass spectrometer tube
Operation (1) Fully automatic startup (including sensitivity calibration) (*1)
(2) Automatic valve operation
(3) Calibration leak automatic temperature compensation
(4) Quick startup function
(5) Zero tracking function
(6) Self-check function
Leak rate display Touch-panel display
: Digital value readout display (Japanese kanji or English can be selected.)
: Analog display (chart/graph display)
Leak rate alarm Audio alarm is provided standard.
I/O RS-232C
Printer (Centronics interface)
Recorder (Linear : 0 to 10 VDC with range output; Analog : 0 to 10 VDC)
External control I/O (device start/stop, test start/end, set points, etc.)
Power supply AC100V 150W standard
Safety standard CE
External dimensions, weight 260 x 480 x 260mm, 14.5kg

*1 Concentration and flow rate units can't be converted directly.
*2 Depending on the settings, some manual operations may be required.