Leak
Detectors
HELIOT 710 Series Plus One
HELIOT 710 series enables various
highest quality leak tests using 'easy to use' conspicuous
color touch panel display. It is a next generation series
with high level sensitivity that is user and eco-friendly.
Features
- Fully automated functionality - including sensitivity calibration.
- In addition to He, H2 leak detection is now offered. (*1)
- Industry’s only large full color touch panel for easy operation.
- 6 language interface installed: English, German, Korean, Chinese (simplified/traditional) and Japanese.
- Smart Function - system surveillance detects trouble before damage occurs to Heliot 710.
- Optional accessories available to suit all user environments – Hand held pendant / Chart Recorder / Cart / Sniffer Probe /Bombing Chambers/ Console type configuration.
- CE certified / IP 30 certified (*2) / RoHS compliant.
1: Vacuum method only. Not applicable with model 711D1.
2: Applicable only with model 711W1, 711D1 and 714D2.
Applications
- Car industry, e.g. torque converter, gasoline
tank, wheel, suspension and brakes.
- Airconditioners e.g. condensor, evaporator,
compressor and heat exchanger
- Electronics/ semiconductors e.g. SAW device,
quartz crystal and IC
- Food/Phramaceutical industry, space develpoment
industry, etc.
- For measuring flow rate, transmittance, adsorption
and sealing etc.
Specifications:
| Model |
711W1 |
714D2 |
| Detectable gas |
Helium, Hydrogen |
| Minimum detectable leak rate |
0.1 × 10-12Pa・m3/sec (Vacuum method/Helium) |
| Analyzer |
90 degrees magnetic field deflection mass spectrometer |
| Main pump |
Turbo molecular pump |
| Fore pump |
Oil rotary pump |
Dry scroll pump |
| Test port |
NW25 |
| Startup time/ Shutdown time |
3min (quick start setting)/3min |
| Size |
480 × 335 × 505mm |
480 × 335 × 600mm |
| Mass |
Approx. 43kg |
Approx. 50kg |
| Input power supply |
AC100V, 110V, 115V, 200V, 220V |
AC100-120V,
AC200-240V |
| Display |
7.5inch TFT color touch panel |
| Test screen |
4type mode
(Graph, meter, value and vacumm system diagram) |
| Input/ Output signal |
Leak rate output Analog style (DC0-10V)/ Printer output/ RS-232C/ Startup, shutdown,setpoint(1,2,3)etc.,for a total of 20 connections |
| Calibration |
Auto calibration during start-up/ External calibration (external calibration leak is an option) |
Download Heliot 710 Brochure.
For
more information |
|
TEL
: 978-686-7550/ FAX : 978-689-6300 |
|